Phone:+86 18710185982
您的当前位置:
BJP-7100Dplus Veterinary Anesthesia Machine
    Publish time 2025-09-19 20:10    
BJP-7100Dplus Veterinary Anesthesia Machine

Anesthesia Machine
Gas Supply:
Gas Source: O2,or O2&Air(Option), or O2&N2O(Option);                         
Gas Source Pressure Gauge: O2,or O2&Air(Option), or O2&N2O(Option);
Pressure of Pipeline Inlet:                                                 
O2, 280kPa~600kPa/41Psi~87Psi, less than 180L/min;
N2O(Option), 280kPa~600kPa/41Psi~87Psi, less than180L/min;
Air(Option), 280kPa~600kPa/41Psi~87Psi, less than 180L/min;
Security Valve: release pressure less than 10 kpa;
O2 Lost Alarm(Option): sound alarm, alimumun cylinder with whistle;
N2O cut off valve protection(Option): when O2 is lost, N2O will be cut off.
Flow Meter:
2 tubes high precision flow meter, 0~1L/min,1.1L/min~10L/min; or 0L/min~10L/min(O2&Air or O2&N2O);
Precision: At 20 degree, atmospheric pressure is 101.3kPa, the precision of flow meter is level:2.5;
O2&N2O Linkage Device(O2&N2O,Option): make sure O2 concentration is not lower than 25%;
O2 Flush: 35L/min~75L/min;
One Key Switch: switch between rebreath and Non-Rebreath;
ACGO Port: non-Rebreath;
Selectatec Mounting Bar: single place used for vaporizer, it is easy to replace;
Vaporizer:
Anesthesia Vaporizer: one of Isoflurane/Halothane/Sevoflurane;
Respiration System/CO2 Circle Absorber/Rebreath:
Soda Lime Chamber: 800ml or 1600ml(Option);                                              
Pop Off Valve: with one key to close; APL valve(Option): 0.19~6kPa;
Airway Pressure Gauge: -20-100cmH2O;

Anesthesia Ventilator
Drive Gas: O2 or Air, Inlet pressure: 280-600 kPa. Max flow≤ 120 L / min;
Safety Pressure: system pressure does not exceed 10 KPa;
Display: 10.1" TFT, with touch screen;
Breath Mode: VCV, PCV, SIMV-VC, SIMV-PC, PSV,Manual/SPONT, SIMV-PRVC (option);
Waveform: Pressure, Flow rate,Volume, Loop, PV loop, PF loop, FV loop
O2 Concentration Monitor: 15%-100%(option);
Electric PEEP: OFF, 3 ~ 30 cmH2O;
Battery Type: Li-ion,work time more than 3 hours;
Parameters Setting Range:                       
Tidal Volume: 50 - 1500ml;                             
Option: 10 - 1500ml;                                    
Respiratory Rate: 1 ~ 100 bpm;                       
Inspiratory Time: 0.1~10.0s;                                                 
Respiratory Rate: 0 ~ 100 bpm;
Percentage of Inspiratory Pause: 0 to 60%;   
PEEP: OFF, 3 ~ 30 cmH2O;                          
Pressure Support: 0 ~ 70 cmH2O;                 
Pressure Control: 5 ~ 70 cmH2O;                  
Flow Trigger: 0.5 ~ 20L/min;                           
Pressure Trigger: 0 ~ 20 cmH2O;                 
PSV Insp Termination Level: 5 ~ 80%;         
 
Monitored Parameters:
Inspiratory Tidal Volume: 0 ~ 3000 mL;
Expiratory Tidal Volume: 0 ~ 3000 mL;
Spontaneous Minute Ventilation: 0 ~ 60 L / min;
Minute Ventilation: 0 ~ 60 L / min;
Respiratory Ratio: 4:1 to 1:10;
Spontaneous Breathing Frequency: 0 ~ 100 bpm;
Compliance: 0 ~ 300 mL/cmH2O;
Respiratory Ratio: 30:1 to 1:150;
Peak Airway Pressure: 0 ~ 100 cmH2O;
Mean Airway Pressure: 0 ~ 100 cmH2O;
PEEP: 0 ~ 100 cmH2O;
Inspiratory Plateau Pressure: 0 ~ 100 cmH2O;
FiO2(option): 15 to 100%;
EtCo2, FiCO2(option);
EtCO2, FiCO2, EtN2O, FiN2O, FiO2, EtO2, EtAA, FiAA, MAC(option);
 
Alarm Settings:                                       
Tidal volume high/low alarm; Minute ventilation high/low alarm; Respiratory rate high/low alarm; Airway pressure high/low alarm; Continuous Positive Airway Pressure/Negative pressure alarm; Apnea alarm; Oxygen deficiency alarm; AC power failure alarm, low battery alarm.
 
Standard Configuration:
Anesthesia machine, one unit X60 Anesthesia vaporizer, one set adult disposable breath circuit, one set pediatric disposable breath circuit, 0.5,1.2,3L airbag, O2, or O2&N2O, or O2&Air gas source pipeline
Option: SIMV-PRVC, O2 Concentration Monitor, O2&N2O Pin Yoke system, EtCo2, FiCO2, EtCO2, FiCO2, EtN2O, FiN2O, FiO2, EtO2, EtAA, FiAA, MAC;
 
Dimension & Weight:
Trolley Stand: 55cm*56cm*153cm, 50kg;