Phone:+86 18710185982
您的当前位置:
BJP-MV 3002
    Publish time 2025-09-27 20:37    

BJP-MV 3002 Cel50 Marked MedicalBreathingApparatus Ventilator

Component Details

Display: 10.4"LcD Display Following Parameters & Real timewaveform display

(1} Gas drive and electrical control, time-pressure switchingand pressure limit control.(2}A high-brightness digital display is used to present thecontrol frequency, tidal volume, throughput, overallrespiratory rate,spontaneous breathing frequency, etc.(3}A highly sensitive and responsive pressure sensor and afloww sensor are used to measure, control and display theairway pressure and gas flow rate and the ventilator isequipped with automatic throughput compensation.(4}In case of an abnormality to the ventilator ormisoperation,the ventilator can raise a visual-audible alarmto automatically protect itself.


BJP-MV 3002

Product Description


Prospective service and support

The HV-600A ventilator is an electrically controlled pneumatic ventilator integrating such functions as time, volume cycling, pressure limit, etc. it provides an alternation in case of irreversible lesions in respiratory muscles or irreversible damage to upper airway to maintain the respiratory function of the patient, and also provides ventilation assistance for the patient during the recovery from a disease or operation.

 

The automatic conversion for standby power while AC power outage, keep working not less than 1 hour.

 

Ventilator Construction:

-Environmental Protection Material construction, hardly to be broken.

-Trolly: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions

 



TECHNIQUES SPECIFICATIONS


Physical specifications


Display

10.4'' LCD Display


Mode

pneumatically powered and electrically controlled system


Trolley

Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions


Technical Specifications


Gas requirement

Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpa


safety valve

<12.5  kPa


Respiratory Rate

1 ~ 99 bpm


Maximum Minute ventilation

≥18L / min


modes of ventilation

A/C, SIPPV,IPPV,SIMV,PSIMV,CPAP,PCV,PEEP,SIGN,SPONT, MANUAL


Ventilator compliance

≤30 Ml/kPa


PEEP

0 ~ 1.0 kPa


Ptr

-1.0kPa~1.0kPa 


IMV frequency

1~20 bpm


I/E Ratio

4:1~1:4, Has the inverse ratio ventilation


Tidal Volume range

0~2000ml


Inspiratory Plateau

0~1s


Maximum Safety  Pressure

≤ 6.0kPa


Sigh

one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point


Pressure limit range

1.0~ 6.0kPa


Airway pressure alarm: Audible and visual and with yellow and red color indicating

Lower: 0.1 kPa ~ 6.0 kPa;   Upper: 0 ~ 5.0  kPa


±0.2 kPa


Ventilation Volume alarm

≥ 18L/ MIN


Power Supply Alarm

Ac/dc power supply are after failing to send out alarm immediately

Alarm Time: keep >120s


Operating conditions


Ambient temperature

10 ~ 40oC


Relative humidity:

no higher than 80 %


Atmospheric pressure

860 hPa ~ 1060 hPa


Power requirement

100-120 Vac, 50/60 Hz; 


Attention: the AC power supply used for the ventilator machine must be well   grounded.


Attention: the ventilator machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993, an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of Anesthesia System.


Storage


Ambient temperature

-15oC ~ +50oC


Relative humidity

no higher than 95 %


Atmospheric pressure

86 kPa ~ 106 kPa.


It should be stored in a room without corrosive gas and well ventilated


Package


packaging box

comply with the requirement of GB/T 15464


Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation


Moisture protection and rain protection to ensure that product is protected from natural damage.


Safty & Alarm


Gas source pressure

Immediate alarm; "high level alarm" indicator flickers raising an audible alarm.


Airway Pressure Alarm

The airway pressure exceeds 15hpa. The maximum delay is no longer than 17s. The "high level alarm" indicator flickers raising an audible alarm.


Tidal Volume Alarm

Lower than 50ml, The alarm will be raised after one breathing cycle.


Ventilation Volume Alarm

Lower: 0~12L/Min;  

Up: 18L/ Min


Battery Voltage Alarm

It alrms when the battery voltage is lower than 10.5V.


Power On Self Test

Self test before working, assure it's working normally.

 



STANDARD CONFIGURATIONS


QTY

NAME


1 set

Main unit


4 pcs

Adult threaded pipe (Silica)


1 set

Child threaded pipe (single use)


1 pc

Oxygen Tube


1pc

Y-Pipe


2pcs

Mask (Silicom)


2pcs

Leather Bag


2pcs

Fuse


1pc

Power Line


1pc

Oxygen pressure reducer


4 pcs

Hoop


1pc

Battery


1pc

Tools Attachment


1 set

humidifier


1 set

English Version User Manual